GB5169
IEC60695-2-10
UL74A
DIN695
| Model | UP-5019 |
| Working voltage | 220V /50Hz or 110V/60Hz (can be customized according client's requirements) |
| Control operation mode | Button control +with infrared remote controller |
| Display | Meters |
| Glow-wire element | Ф4mm ± 0.04mm (> 77 % Ni/20 ± 1 % Cr) special standard ring shape, horizontal standing |
| Specimen pressure on glow-wire element | 0.95N±0.1N |
| Maximum depth of heating | 7mm ± 0.5mm |
| Specimen movement speed | 10mm /s ~ 25mm /s |
| Temperature calibration | Calibration temperature:960°±10°C The glow wire element is heated by electric heating |
| Thermocouple: imported original BB 1mm armour nickel-chromium/nickel/ nickel aluminum K-Thermocouple, | |
| Temperature resistance up to 1100°C | |
| Silver foil: purity 99.8%, area :2mm², thickness:0.06mm | |
| Tray | A layer of silk paper (12~30g/m³) is covered on the 10mm thick pine board, with a distance of 200±5mm below the test sample |
| Test chamber | >0.5 cube, sample distance from the inside of the chamber is greater than 100mm, black yogon background,background illuminance ≤ 20Lx,with exhaust fan |
| Glow-wire time | 30s±1s |
| Test procedure | The test sample is clamped on the movable fixture, the movable fixture can run forward automatically after testing started and will return automatically after glow-wire time is finished |
| Lighting | Equipped with lighting in the test chamber |
| Dimension | 1100*550*1200mm, 140kg |
Our service:
During the whole business process, we offer Consultative Selling service.
FAQ:
Moreover, If your machine doesn’t work, you can send us an e-mail or call us we will try our best to find the problem through our conversation or by video chat if necessary. Once we have confirmed the problem, the solution will be offered within 24 to 48 hours.